Statox 560 gas detector – self-test with real gas
Detection of Phosgene COCl2, Chlorine Cl2, Hydrogen Cyanide HCN, Carbon Monoxide CO, Hydrochloric Acid HCl, Ammonia NH3 or Hydrogen Sulfide H2S (others upon request) at the highest safety level!
Functional safety with SIL2 is a measure of the extremely low probability of failure if required. In addition, a self-test with real gas guarantees that a fault – no matter how unlikely – will not go unnoticed with this gas warning device.
The self-test can also be used to determine the need for maintenance. This saves a lot of time for inspection rounds.
Quick navigation:
Features | Technical data | Self test | Downloads
Features of the Statox 560 gas detector
The Statox 560 is a state-of-the-art product in every respect:
- The new generation of sensors impresses with its high sensitivity, very fast response, low cross-sensitivity and plug-and-play technology.
- The gas generators produce the gas to be measured during the self-test. This ensures that the sensor functions perfectly in terms of measurement technology.
- The Statox 560 is a 4 – 20 mA transmitter.
- The entire electronics are housed in an Ex d enclosure with a window. The self-test takes place automatically once a day at a programmable time. However, it can also be triggered manually on site using a magnetic pen. The self-test can also be started from your control room. This is done with a separate button or with the Statox 503 Control Module.
- The magnetic pen supplied can be used to set parameters and call up various information. As the housing does not need to be opened, no fire licence is required.
- The sensor module contains the sensor, the generator cell and a micropump. It is designed in protection class Ex i. A fire permit is therefore not required to work on this module. The module can be completely removed and replaced with a new one. This design reduces the work required in the field to an absolute minimum.
More safety and more user-friendliness is not possible!
Technical data gas detector Statox 560
Product Name: | Statox 560 (Type 5377) |
Power Supply: | 24 (16-30) VDC |
Power Consumption: | max. 2.7 W (8.7 W for COCl2) at an input voltage ≤ 26 VDC |
Operating Temperature: | -30 to +60 °C |
Pressure Range: | 700 to 1300 hPa |
Humidity Range: | 0% to 99% r. H., non-condensing |
Application Range: | II 2G |
Explosion Protection: | Ex d ib IIC T4 Gb (Um = 30 VDC for all connections!) |
EU Type Approval Certificate: | BVS 16 ATEX E 065 X |
Ingress Protection: | IP66 (Gas entry opening IP54) |
Display: | 2 x 16 characters, illuminated |
Housing: | Aluminium die-cast with epoxy coating / stainless steel |
Connections: | 10-wire cable tail (1 m) or 7-pole eXLink plug connection |
Open-Drain Outputs: | 2 x Alarm, 1 x System Error, 1 x Maintenance Required Connection ratings: each max. 30 VDC / 2.5 A |
SF-Open-Drain Output: | Active in normal operation (conducting) |
Analog Output: | 0 mA in case of error; 2 or 4 mA in service mode, adjustable 4 – 20 mA in measurement operation 22 mA for overrange max. load: 450 Ohm |
EMC: | EN 61000-6-4:2007 + A1:2011 / EN 50270:2015 (Type 2) |
Functional Safety: | SIL 2 capable according to IEC 61508:2010 |
Automatic Self-Test: | every 24 hours, selectable time |
Weight: | approx. 4800 g |
Dimensions (WxHxD): | 121 x 294 x 138 mm |
Self-test
Gas | Formula | Test Gas |
Chlorine | Cl2 | Cl2 |
Carbon Monoxide | CO | H2 |
Phosgene | COCl2 | COCl2 |
Hydrogen Sulfide | H2S | H2S |
Hydrogen Cyanide | HCN | HCN |
Ammonia | NH3 | H2S |
Hydrogen Chloride | HCl | H2S |