Statox 560 gas detector – self-test with real gas

Detection of Phosgene COCl2, Chlorine Cl2, Hydrogen Cyanide HCN, Carbon Monoxide CO, Hydrochloric Acid HCl, Ammonia NH3 or Hydrogen Sulfide H2S (others upon request) at the highest safety level!

Functional safety with SIL2 is a measure of the extremely low probability of failure if required. In addition, a self-test with real gas guarantees that a fault – no matter how unlikely – will not go unnoticed with this gas warning device.

The self-test can also be used to determine the need for maintenance. This saves a lot of time for inspection rounds.

Quick navigation:
Features | Technical data | Self test | Downloads

Features of the Statox 560 gas detector

Gaswarngerät Statox 560

The Statox 560 is a state-of-the-art product in every respect:

  • The new generation of sensors impresses with its high sensitivity, very fast response, low cross-sensitivity and plug-and-play technology.
  • The gas generators produce the gas to be measured during the self-test. This ensures that the sensor functions perfectly in terms of measurement technology.
  • The Statox 560 is a 4 – 20 mA transmitter.
  • The entire electronics are housed in an Ex d enclosure with a window. The self-test takes place automatically once a day at a programmable time. However, it can also be triggered manually on site using a magnetic pen. The self-test can also be started from your control room. This is done with a separate button or with the Statox 503 Control Module.
  • The magnetic pen supplied can be used to set parameters and call up various information. As the housing does not need to be opened, no fire licence is required.
  • The sensor module contains the sensor, the generator cell and a micropump. It is designed in protection class Ex i. A fire permit is therefore not required to work on this module. The module can be completely removed and replaced with a new one. This design reduces the work required in the field to an absolute minimum.

More safety and more user-friendliness is not possible!

Contact us

Technical data gas detector Statox 560

Product Name: Statox 560 (Type 5377)
Power Supply: 24 (16-30) VDC
Power Consumption: max. 2.7 W (8.7 W for COCl2) at an input voltage ≤ 26 VDC
Operating Temperature: -30 to +60 °C
Pressure Range: 700 to 1300 hPa
Humidity Range: 0% to 99% r. H., non-condensing
Application Range: II 2G
Explosion Protection: Ex d ib IIC T4 Gb (Um = 30 VDC for all connections!)
EU Type Approval Certificate: BVS 16 ATEX E 065 X
Ingress Protection: IP66 (Gas entry opening IP54)
Display: 2 x 16 characters, illuminated
Housing: Aluminium die-cast with epoxy coating / stainless steel
Connections: 10-wire cable tail (1 m) or 7-pole eXLink plug connection
Open-Drain Outputs: 2 x Alarm, 1 x System Error, 1 x Maintenance Required
Connection ratings: each max. 30 VDC / 2.5 A
SF-Open-Drain Output: Active in normal operation (conducting)
Analog Output: 0 mA in case of error;
2 or 4 mA in service mode, adjustable
4 – 20 mA in measurement operation
22 mA for overrange
max. load: 450 Ohm
EMC: EN 61000-6-4:2007 + A1:2011 / EN 50270:2015 (Type 2)
Functional Safety: SIL 2 capable according to IEC 61508:2010
Automatic Self-Test: every 24 hours, selectable time
Weight: approx. 4800 g
Dimensions (WxHxD): 121 x 294 x 138 mm
Gaswarngerät Statox 560 - Selbsttest

Self-test

Gas Formula Test Gas
Chlorine Cl2 Cl2
Carbon Monoxide CO H2
Phosgene COCl2 COCl2
Hydrogen Sulfide H2S H2S
Hydrogen Cyanide HCN HCN
Ammonia NH3 H2S
Hydrogen Chloride HCl H2S

Downloads

Gaswarngerät Statox 503

Suitable control module:
Statox 503